Search our site:      Login | Register      

Bluewave Specifications

Measuring Range:

0.01to 2000 microns

Lasers:

Wavelength Red 780nm , Blue 405 nm

Power:

3mW nominal

Detection System:

Two fixed photo-electric detectors with logarithmically spaced segments placed at correct angles for optimal scattered light detection. 0.02 to 163 degrees using 151 detector segments.

Data Handling:

Volume, Number and Area distributions as well as percentile and other summary data. Data is stored in ODBC format in encrypted Microsoft Access Databases to ensure compatibility with external statistical software applications. Data integrity may be ensured using FDA 21 CFR Part 11 compliant security features including password protection, electronic signatures and assignable permissions.

Typical Analysis Time:

10 to 30 seconds

Electrical:

AC input: 90 – 132 VAC, 47 - 63 Hz, single phase
200 to 265 VAC, 47 – 63 Hz, single phase

Power Consumption:

25 VA maximum

Environmental:

Temperature: 10 to 35 Degrees C.

Humidity: 90% RH, non condensing maximum
Storage Temperature:

-10 to 50 Degrees C (Dry only)

Pollution:

Degree 2

Physical Specifications
Finishes:

Case Material Steel and impact resistant plastic
Exterior Surfaces finished with corrosion resistant paint or plating

Dimensions:

14H x 22W x 18D in (360H x 560W x 460D mm)

Weight: 60 lbs (27 kg)
Dry Operation
Eductor Air Supply:

100 psi (689 kPa) maximum pressure
5 CFM at 50 psi (345 kPa) minimum flow rate
Free of dry contaminants, moisture and oil.

Vacuum:

No greater than 50 CFM