Particle Size: 0.25 to 2816 microns
Physical
Dimensions: 7.5"H x 6.5"W x 13.5"D ( 191H x 165W x 343D mm)
Weight: 10 lbs (4.5 kg)
Case Material: Door - Nalgene plastic; Covers - ABS Plastic
Electrical
Power: 24VDC provided by Microtrac Sample Delivery Controller (SDC)
Power Consumption: Included in maximum VA rating of SDC
Recirculator Interface Standard: 9 pin male-male straight through interface cable
Environmental
Temperature (operating): 10 to 35 Degrees C
Temperature (storage): -10 to 50 Degrees C
Humidity: 90% RH, non condensing maximum
Pollution: Degree 2
Connection Ports
Vacuum Connection: Quick disconnect fitting on S3500, 1.25 IN tube (Standard Vacuum hose.) If supplied
by user, vacuum source must exceed 50 CFM, 50 "H2O
Eductor Sample Connection: 3/8" tube fitting
Eductor Air Connections: Push to connect fittings
Eductor Air Supply: 100 psi (689kPa) maximum supply pressure
5 CFM at 50 psi (345 kPa) minimum flow rate
free of dry contaminants, moisture and oil.
Materials in Contact with Sample
Dry operation: Stainless steel, anodized aluminium (black chrome plated, stainless steel optional),
PVC, HDPE
Vacuum (subject to change by manufacturer): paper, cotton, glass fibre, aluminium, EVA