Search our site:      Login | Register      

Turbotrac Specifications

Particle Size: 0.25 to 3000 microns
 

Physical

Dimensions 5.25 H x 5 W x 13.5D in. ( 133H x 127W x 343D mm)
Weight: 9 lbs (4.1 kg)
Case Material: Steel, exterior surfaces finished with corrosion resistant paint or plating.

Electrical

Power: 12VDC and 24VDC provided by Microtrac Sample Delivery Controller (SDC)
Power Consumption: Included in maximum VA rating of SDC
Recirculator Interface Standard: 9 pin male-male straight through interface cable

Environmental

Temperature (operating): 10 to 35 Degrees C
Temperature (storage): -10 to 50 Degrees C
Humidity: 90% RH, non condensing maximum
Pollution: Degree 2

Connection Ports

Vacuum Connection: Quick disconnect fitting on S3500, 1.25 in tube (Standard Vacuum hose.) If supplied
by user, vacuum source must exceed 50 CFM, 50 "H2O
Eductor Sample Connection: 3/8" tube fitting
Eductor Air Connections: Push to connect fittings
Eductor Air Supply: 100 psi (689kPa) maximum supply pressure
5 CFM at 50 psi (345 kPa) minimum flow rate
free of dry contaminants, moisture and oil.

Materials in Contact with Sample

Dry operation: Stainless steel, anodized aluminium (black chrome plated, stainless steel optional),
PVC, HDPE

Vacuum (subject to change by manufacturer), paper, cotton, glass fibre, aluminium, EVA